Case Report: A novel method of needle-free transseptal puncture

Introduction

A successful transseptal puncture (TSP) is essential to gain access to the left atrium (LA) during catheter ablation procedures for atrial fibrillation (AF) (1). TSP has become the standard approach to cross the atrial septum safely. Traditionally, TSP is performed using an SL0 sheath in combination with a Brockenbrough (BRK) transseptal needle (St. Jude Medical, USA) (2). After crossing the septum, operators often exchange the initial 0.032 inch J-tip guidewire for a GMS-1 guidewire (0.025 inch, pigtail configuration; Toray Medical Co., Ltd., Japan) to facilitate device advancement into the LA.

Since the first TSP procedure in the late 1950s (36), numerous technical advancements have been integrated into clinical practice (711), including the adoption of zero-fluoroscopic TSP under intracardiac echocardiography (ICE) guidance (12). However, in our practice, we identified several drawbacks associated with performing fluoroless TSP using either the BRK needle or electrocautery guidewire. The BRK needle often falls out of the echocardiographic frame, increasing the risk of atrial wall perforation due to poor visualization of the needle tip. Additionally, the J-tip guidewire of the SL0 sheath often provides suboptimal support, complicating septal crossing with the SL0 sheath.

In response to these challenges, we explored the feasibility of performing TSP directly using a 0.025 inch GMS-1 guidewire (Figure 1). This approach may offer a more efficient and safer alternative to traditional methods.

www.frontiersin.org

Figure 1. GMS-1 guidewire-based electrocautery TSP. (A) SL0 sheath and a GMS-1 guidewire. (B) Preloaded before TSP. (C) The tip of the guidewire formed a pigtail configuration after crossing.

Case report Case 1

A 61-year-old male was referred to our hospital for radiofrequency pulmonary vein isolation (PVI) ablation due to highly symptomatic, drug-refractory paroxysmal atrial fibrillation (AF). Additionally, a computed tomography (CT) scan was performed to merge morphological and electroanatomic information during the ablation.

The procedure was performed under intravenous general anesthesia. An intracardiac echocardiography (ICE) catheter (SoundStar, Biosense Webster) was introduced via the left femoral vein, providing optimal imaging of the atrium and interatrial septum. After creating an anatomical map with real-time ICE and CARTO-Sound Image Integration Module, a merged CT and Sound map was acquired. We placed a 10-pole catheter into the coronary sinus under CARTO and ICE guidance. Subsequently, a 0.032 inch guidewire was advanced from the right femoral vein to the superior vena cava (SVC) to guide the transseptal sheath (SL0, St. Jude Medical). Once the sheath reached the SVC, the 0.032 inch guidewire was exchanged for a GMS-1 guidewire (0.025 inch, pigtail-shaped). Under real-time visualization, the transseptal sheath (SL0, St. Jude Medical) was introduced into the right atrium (RA).

Under ICE guidance, the sheath and the GMS-1 wire were positioned against the septum. This was confirmed by the typical “tenting” appearance on echocardiographic imaging when slight pressure was applied to the septal wall (Figure 2A). The wire was then advanced through the septum into the left atrium (LA) using electrocautery technique with an energy of 30–35 W for less than 1 s under “dry cut” mode by ERBE VIO 10°C generator (ERBE Elektromedizin GmbH, Germany). Upon accessing the LA, the wire's tip immediately assumed a pigtail configuration, preventing further tissue penetration (Figure 2B; Supplementary Video S1). Subsequently, the SL0 sheath and its dilator were smoothly introduced into the LA. A second TSP with Vizigo (Biosense Webster) was done smoothly using the same technique. PVI was subsequently completed smoothly.

www.frontiersin.org

Figure 2. Transseptal process under ICE. (A) Identify the septum by typical tenting image. (B) The GMS-1 guidewire formed a pigtail shaped. For a complete video clip, please refer to Supplementary Video S1.

Case 2

A 69-year-old female presented to our center for radiofrequency ablation with PVI due to symptomatic persistent AF. Pre-procedural TEE confirmed the absence of thrombi in the LA appendage. As in the previous case, both TEE and a CT scan were conducted pre-procedurally.

The procedure was conducted under intravenous general anesthesia. A 10-pole catheter was first introduced into the coronary sinus. Subsequently, a 0.032 inch guidewire was advanced from the right femoral vein to the SVC to guide the transseptal sheath (SL0, St. Jude Medical). Once the sheath reached the superior vena cava (SVC), the 0.032 inch guidewire was exchanged for a GMS-1 guidewire (0.025 inch, pigtail-shaped).

Under fluoroscopic guidance, the interatrial septum was identified, and the tip of the guidewire was positioned against the septum. TSP was then successfully performed using the electrocautery technique similar as above (Figure 3; Supplementary Video S2), allowing the smooth introduction of the SL0 sheath and its dilator into the LA. A second puncture with Agilis NxT Steerable Introducer (Abbott Medical Australia Pty Ltd) was performed using the same technique. The catheter ablation was subsequently completed without complications.

www.frontiersin.org

Figure 3. Transseptal process under fluoroscopy. The GMS-1 guidewire became pigtail configuration after crossing the septum. For a complete video clip, please refer to Supplementary Video S2.

Discussion

In this case report, we demonstrate a novel electrocautery-assisted TSP technique using the cost-effective and widely accessible GMS-1 guidewire. This approach not only enhances the safety of TSP by providing additional support when advancing the sheath and reducing the need for guidewire exchanges within the LA, but also improves procedural efficiency. This technique is particularly advantageous when employing two-dimensional ICE to pursue a zero-fluoroscopy procedure.

Electrocautery-assisted TSP using guidewires has become increasingly utilized among experienced operators in modern era (7, 8, 10, 12, 13). Compared with previous studies, our approach offers several key advantages:

1. Safety: Unlike TSP using traditional mechanical needles, the GMS-1 guidewire immediately assumes a pigtail configuration upon crossing the septum, which minimizes the risk of atrial wall perforation (7, 8). This is particularly beneficial during second TSP with ICE guidance, because the image under ICE was often interfered by the first transseptal sheath.

2. Efficiency: The absence of a need for guidewire exchange streamlines the procedure, reducing the time required and mitigating the risk of thrombus formation or air bubble introduction.

3. Support for Large Sheath Exchange: The GMS-1 guidewire is designed to facilitate the smooth delivery of large sheaths, such as the SL0 sheath or steerable sheath, through the septum, enhancing procedural efficiency and ease of use.

4. Cost-effective: Although several commercial kits have demonstrated the efficacy and safety in needle-free and fluoroless TSP, their cost still remains a significant concern (7, 11). Our technique offers a more affordable alternative with comparable efficacy.

With the rising popularity of single-shot procedures, including cryoablation and pulsed field ablation, larger sheaths are increasingly required (1416). In our experience, we have successfully delivered a cryoablation sheath (Medtronic, FlexCath Advance™ steerable sheath, 15 Fr) using the GMS-1 guidewire, suggesting that our technique could support these larger sheaths effectively. Further research and case studies are warranted to evaluate the technique's broader applicability in this context.

In conclusion, we presented two cases demonstrating the safety, efficacy, and cost-effectiveness of needle-free and zero-fluoroscopic electrosurgery-assisted TSP using a 0.025 inch GMS-1 guidewire. This approach appears to be a viable and economical alternative to traditional methods.

Data availability statement

The original contributions presented in the study are included in the article/Supplementary Material, further inquiries can be directed to the corresponding author.

Ethics statement

Written informed consent was obtained from the individual(s) for the publication of any potentially identifiable images or data included in this article.

Author contributions

C-CL: Writing – original draft, Writing – review & editing. C-AC: Writing – review & editing. S-HC: Writing – review & editing. C-KC: Writing – review & editing. Y-SC: Writing – review & editing. C-EL: Writing – review & editing. T-PJ: Writing – review & editing. C-CY: Conceptualization, Funding acquisition, Writing – review & editing.

Funding

The author(s) declare that no financial support was received for the research, authorship, and/or publication of this article.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher's note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Supplementary material

The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fcvm.2024.1493240/full#supplementary-material

Supplementary Video S1 | A video clip showed the moment of atrial septal crossing under fluoroscopy. The guidewire curved immediately after a successful TSP.

Supplementary Video S2 | A video clip showed the moment of atrial septal crossing under ICE. TSP, transseptal puncture; ICE, intracardiac echocardiography.

References

2. Russo G, Taramasso M, Maisano F. Transseptal puncture: procedural guidance, challenging situations and management of complications. EuroIntervention. (2021) 17:720–7. doi: 10.4244/EIJ-D-20-00454

PubMed Abstract | Crossref Full Text | Google Scholar

4. Brockenbrough EG, Braunwald E. A new technic for left ventricular angiocardiography and transseptal left heart catheterization. Am J Cardiol. (1960) 6:1062–4. doi: 10.1016/0002-9149(60)90361-1

Crossref Full Text | Google Scholar

5. Ross J Jr, Braunwald E, Morrow AG. Transseptal left atrial puncture; new technique for the measurement of left atrial pressure in man. Am J Cardiol. (1959) 3:653–5. doi: 10.1016/0002-9149(59)90347-9

PubMed Abstract | Crossref Full Text | Google Scholar

7. de Asmundis C, Chierchia GB, Sarkozy A, Paparella G, Roos M, Capulzini L, et al. Novel trans-septal approach using a safe sept J-shaped guidewire in difficult left atrial access during atrial fibrillation ablation. Europace. (2009) 11(5):657–9. doi: 10.1093/europace/eup089

PubMed Abstract | Crossref Full Text | Google Scholar

8. Khan JM, Rogers T, Eng MH, Lederman RJ, Greenbaum AB. Guidewire electrosurgery-assisted trans-septal puncture. Catheter Cardiovasc Interv. (2018) 91:1164–70. doi: 10.1002/ccd.27311

PubMed Abstract | Crossref Full Text | Google Scholar

9. Luani B, Zrenner B, Basho M, Genz C, Rauwolf T, Tanev I, et al. Zero-fluoroscopy cryothermal ablation of atrioventricular nodal re-entry tachycardia guided by endovascular and endocardial catheter visualization using intracardiac echocardiography (ice&ICE trial). J Cardiovasc Electrophysiol. (2018) 29(1):160–6. doi: 10.1111/jce.13354

PubMed Abstract | Crossref Full Text | Google Scholar

10. Razminia M, Willoughby MC, Demo H, Keshmiri H, Wang T, D'Silva OJ, et al. Fluoroless catheter ablation of cardiac arrhythmias: a 5-year experience. Pacing Clin Electrophysiol. (2017) 40(4):425–33. doi: 10.1111/pace.13038

PubMed Abstract | Crossref Full Text | Google Scholar

11. Rizzi S, Pannone L, Monaco C, Bisignani A, Miraglia V, Gauthey A, et al. First experience with a transseptal puncture using a novel transseptal crossing device with integrated dilator and needle. J Interv Card Electrophysiol. (2022) 65(3):731–7. doi: 10.1007/s10840-022-01329-3

PubMed Abstract | Crossref Full Text | Google Scholar

12. Cheng H, Bangash AB, Tao Y, Jiang R, Liu Q, Cheng C, et al. Intracardiac echocardiography guided electrified J-wire trans-septal puncture: a prospective randomized controlled trial. Pacing Clin Electrophysiol. (2024) 47(3):448–54. doi: 10.1111/pace.14916

PubMed Abstract | Crossref Full Text | Google Scholar

13. Wasserlauf J, Knight BP. Comparing the safety and effectiveness of dedicated radiofrequency transseptal wires to electrified metal guidewires. J Cardiovasc Electrophysiol. (2022) 33:371–9. doi: 10.1111/jce.15341

PubMed Abstract | Crossref Full Text | Google Scholar

14. Andrade JG, Wells GA, Deyell MW, Bennett M, Essebag V, Champagne J, et al. Cryoablation or drug therapy for initial treatment of atrial fibrillation. N Engl J Med. (2021) 384(4):305–15. doi: 10.1056/NEJMoa2029980

PubMed Abstract | Crossref Full Text | Google Scholar

15. Reddy VY, Gerstenfeld EP, Natale A, Whang W, Cuoco FA, Patel C, et al. Pulsed field or conventional thermal ablation for paroxysmal atrial fibrillation. N Engl J Med. (2023) 389(18):1660–71. doi: 10.1056/NEJMoa2307291

PubMed Abstract | Crossref Full Text | Google Scholar

16. Wazni OM, Dandamudi G, Sood N, Hoyt R, Tyler J, Durrani S, et al. Cryoballoon ablation as initial therapy for atrial fibrillation. N Engl J Med. (2021) 384(4):316–24. doi: 10.1056/NEJMoa2029554

PubMed Abstract | Crossref Full Text | Google Scholar

留言 (0)

沒有登入
gif