Very Early Loading of a Bone-Anchored Hearing System Is Safe

Objective 

To investigate the feasibility of early sound processor loading with two separate bone-anchored hearing aid implants 1 to 2 weeks after implantation. Secondary endpoints included registration of any postoperative complications, skin reactions, and any adverse events.

Study Design 

Prospective, nonrandomized, noncontrolled, multicenter clinical trial.

Setting 

Tertiary referral center.

Patients 

Sixty adult patients, eligible for bone-anchored hearing aid surgery, were enrolled.

Intervention 

Implantation of the bone-anchored hearing aid implants were done using minimally invasive surgery with tissue preservation.

Main Outcome Measure 

Implant stability quotient (ISQ) values were recorded using resonance frequency analysis. Follow-ups were scheduled 5 to 12 days, 4 to 8 weeks, 6 months, and 12 months postoperatively. Differences between the lowest ISQ values (ISQ low) over time were analyzed by paired-samples t test with a significance level of 0.05. Skin and soft tissue reactions were assessed according to Holger's classification.

Results 

Loading time of the Ponto Wide implant was 11.6 ± 1.4 days (SD) in average (median, 12 d; range, 7–19 d). Loading of the Ponto BHX implant was done 9.7 ± 3.5 days (SD) postoperatively (median, 9 d; range, 5–19 d). The implant stability increased significantly during the 12-month study period with both implants. One Ponto BHX implant was lost. Skin and soft tissue reactions were few in both groups. Across all planned follow-up visits, no skin and soft tissue reactions were observed in 86% (Ponto BHX implant) and 85% (Ponto Wide implant).

Conclusion 

Very early loading of sound processors after percutaneous bone-anchored hearing system surgery is safe already 1 to 2 weeks postoperatively.