To our knowledge, the literature lacks information regarding PLGA, or chitosan augmented with meropenem, to enhance the titanium plates used in oral and maxillofacial surgery. As a consequence, the present research was undertaken to verify and investigate the application and efficacy of PLGA, or CH enhanced with meropenem as a coating for titanium plates to prevent bacterial adhesion and colonization. As a prophylactic against infection in routine maxillofacial surgeries, particularly in reconstruction cases when the patient may have numerous comorbidities, these redesigned titanium plates are intended to replace their present counterparts. Moreover, this might reduce the chances of subjecting the patients to further surgeries and might reduce the burden upon the health care system.
In this study, the ultrastructural topography of coated titanium plates before and after exposure to two types of bacteria (Gram + ve Staphylococcus aureus [SA] and Gram -ve Pseudomonas aeruginosa [PA]) were evaluated using a scanning electron microscope to study the copolymer layer characteristics over the surface of the plate with or without bacterial adherence.
2. Materials and Methods 2.1. Sample PreparationThe present study was carried out on 54 single-hole segments of 0.4 mm thick, low-profile MatrixMIDFACE titanium plates (DePuy Synthes, Warsaw, IN, USA).
Coating materials were divided into two categories:
A.Carriers: composed of copolymers PLGA and CH
B.Antimicrobial agent: MEM
The following coating materials—PLGA or CH and MEM—were investigated using 1% w/v in acetone (Sigma-Aldrich, St. Louis, MI, USA) of each solution. They were sprayed onto the entire surface of the plates using an airbrush system (Harder & Steenbeck, Oststeinbek, Germany) [12]. The plates were left to air-dry overnight at room temperature in the safety cabinet. The coating procedure was carried out accordingly:Solution 1: 25 mg PLGA and 5 mg meropenem in 2.5 mL acetone (MP)
Solution 2: 25 mg chitosan and 5 mg meropenem in 2.5 mL acetone (MC)
Solution 3: 5 mg meropenem in 2.5 mL acetone (MEM)
Solution 4: 25 mg PLGA in 2.5 mL acetone (PLGA)
Solution 5: 25 mg chitosan in 2.5 mL acetone (CH)
Each plate was weighed before and after spraying to ensure the final modified plates contained 5 mg of MEM and 25 mg of the polymer [13]. 2.2. Bacteria StrainsGram-positive aerobic (Staphylococcus aureus “SA” ATCC® # 25923) and Gram-negative aerobic (Pseudomonas aeruginosa “PA” ATCC®# 27853) bacterial strains were selected for this research. Bacterial adhesion, proliferation, and growth, along with the biofilm formation, were studied accordingly.
2.3. Bacteria CultivationTryptic soy broth (TSB; Sigma Aldrich, St. Louis, MI, USA) was used to cultivate the two types of bacteria (SA and PA). The cultivation took place overnight in an incubator set at 37 °C. The following day, the culture was diluted to 1:50 in fresh TSB and incubated for 3 h at 37 °C to achieve the logarithmic growth phase. A total of 25 μL (equal to 50 McFarland) of the bacteria culture was injected into 18 wells on six lines of a flat-bottomed 96-well microtiter polystyrene plate. One plate was used for each type of bacteria. One segment of each coated plate was placed into the 18 wells with the bacteria solution, including the uncoated negative control group. The polystyrene plate was gently spun at 20 rpm for 24 h and kept at 37 °C to generate biofilms [14,15]. 2.4. Sample Preparation for Scanning Electron Microscope “SEM”Each titanium plate segment was removed from the polystyrene plate using sterile forceps and gently washed with phosphate buffered saline (PBS; Sigma Aldrich, St. Louis, MI, USA) to remove planktonic bacterial cells. The plates were immersed for 4 h in 25% glutaraldehyde solution (Sigma Aldrich, St. Louis, MI, USA), then rinsed three times with PBS. The plates were dehydrated in ethanol solutions of increasing concentrations (50, 70, 80, 90, and 100%) for 5 min at each concentration. Samples were stored in 100% ethanol at 4 °C until examination time [16]. 2.5. SEM ProtocolThe processed samples were affixed to the stubs with double-sided carbon conductivity tape. A thin layer of gold was sprayed onto the samples using a Quorum sputter coater (Q150RS, Laughton, East Sussex, UK) under a vacuum in an argon atmosphere for 1 min. The biofilm adhesion to the titanium plates surface was examined at an accelerating voltage of 10 kV under a vacuum 4 × 106 bar using Zeiss (EVO-LS10, Jena, Germany). The images were captured using Smart SEM software, version 5.05 (Zeiss, Jena, Germany) [16,17]. 2.6. SEM Image AnalysisThe bacteria were counted using ImageJ software (version 1.53, MD, Bethesda, USA), and the plates without bacteria were qualitatively analyzed [18]. 2.7. Statistical AnalysisThe collected data were analyzed using the Statistical Package for the Social Sciences software version 26.0 (IBM Inc., Chicago, IL, USA). Descriptive statistics (mean and standard deviation) were used to express all quantitative variables. Kruskal–Wallis nonparametric test was performed to compare the quantitative data obtained from the SEM imaging of the samples because normality was not satisfied. A Mann–Whitney U test was used to compare within and between the different materials. One examiner carried out all assessments and repeated them three times to confirm reproducibility and reliability. The results were considered statistically significant when p ≤ 0.05.
4. DiscussionBacteria were discovered by Leeuwenhoek over three centuries ago. Since then, scientists have continuously attempted to eliminate existing and emerging infectious diseases caused by bacteria and develop antibacterial agents for treatment [19].To date, the literature lacks information regarding bacterial colonization on titanium microplates used in oral and maxillofacial surgery. However, multiple studies have investigated the microbiological and biofilm characteristics of bacteria on medical devices implanted in the body or on human tissue [20]. The prime cause of orthopedic implant-related infections was attributed to SA. SA was also reported on cardiac implants, vascular grafts, and cosmetic implants [21,22].Infection is caused primarily by bacterial colonies growing in protected biofilms on the titanium surface. Biofilms are multicellular communities bound together by an extracellular matrix produced by bacteria. The techniques various bacteria employ to create biofilms differ based on environmental factors and strain characteristics [19]. The adherent bacterial cells form multicellular aggregates encased in an extracellular polymeric substance-based matrix [23]. As biofilms are formed over the plates, it is difficult for the antibiotics to enter and clear the infection. Hence, antibiotic therapy is of little use in cases of infection associated with osteosynthesis plates. Archer et al. interpreted that those implanted devices’ surfaces become covered with host proteins that provide a surface for bacterial adhesion and biofilm formation. Once the device becomes infected, it must be completely removed because SA biofilm cannot be destroyed by antibiotic treatments [22,24,25]. Biofilms are one of the main factors in chronic infections on the grounds of higher tolerance to antibiotics given that they can combat phagocytosis and the immune system’s other defensive strategies [26]. Therefore, the bacteria within biofilms become resistant to several antibiotics, which drives biofilms to an imminent predicament in therapeutics [27].Nevertheless, antibiotic tolerance is predominantly dependent on the formation of biofilm; the structure of the extracellular matrix with proteins, lipids, water, glycolipids, polysaccharides, and surfactants; and the architecture of the biofilm, which pertains to how the biomass is organized within the biofilm [28,29]. Consequently, the present investigation was undertaken to evaluate the topographical and ultrastructural architecture of titanium plates coated with PLGA, CH, and/or MEM with or without SA or PA bacteria.The overall topographical characteristics of the titanium plate surface showed a relatively uniform and homogeneous coating in each group, with no impact on plate shape or microporosity. The most frequently isolated microorganisms in implant-associated infections are Gram-positive cocci, with SA being the most recurrent cause. Gram-negative bacteria such as PA are responsible for 10–23% of all cases causing infections. PA is considered one of the most difficult-to-treat bacteria because of the growth rate of its multidrug-resistant strains and its ability to develop particular virulence and persistence mechanisms, such as biofilm formation and the production of small colony variants [30].The highest number of bacteria was found in the uncoated plates associated with PA. This is a typical interpretation given the titanium alloy’s lack of innate antibacterial properties and the virulence of PA compared to SA, resulting in the highest bacterial count among all the groups. As for the PLGA- and CH-coated plates, the resulting bacterial count was substantially lower compared to the negative control group.
The coated plates released MEM into their surroundings immediately after placement to avoid bacterial attachment and biofilm development. Successively, the constant release of bactericidal doses took up to 3 days. To avoid bacterial resistance, all the combined drugs must be released over 3–4 days at the desired site to prevent the initial infection. Undeniably, the release of subinhibitory antibiotic concentrations linked with prolonged treatment time is known to cause bacterial resistance to antimicrobial agents [13,31].Within the present findings, bacterial colonization was readily observed on the uncoated titanium plates. The highest SA bacterial inhibition was related to the MP group, and the CH presented the highest inhibition in the PA group. A statistically significant difference was seen between the different materials in the SA and PA groups. Furthermore, a statistically significant difference was noticed in the PLGA-, MP-, and CH-coated plates regarding SA and PA. However, there was no statistically significant difference between the SA and PA groups in relation to the MC- and MEM-coated and uncoated plates.
5. ConclusionsThe modified titanium plates with polymeric carriers (such as PLGA or CH) supplemented with meropenem seemed to enhance the bacterial inhibition compared to the uncoated plates, reducing the biofilm formation, and hindering bacterial proliferation. Accordingly, if a bacterial biofilm accumulates on the plate’s surface, the presence of the antibacterial agent at the plate site can directly increase its antibacterial efficacy.
The surface treatment of titanium plates is a challenging, useful approach to avoid infection risks. Further investigations are recommended to acquire new biofunctionalization approaches to improve the antibacterial efficacy of the titanium plates by modifying the polymer properties and, in turn, the antibiotic release profile.