An audit of the diagnostic accuracy of rotational thromboelastometry for the identification of hypofibrinogenaemia and thrombocytopenia during cardiopulmonary bypass

Department of Anaesthesia, Sir Charles Gairdner Hospital and St John of God Hospital, Perth, Western Australia

We audited the diagnostic accuracy of ROTEM® (TEM Innovations, GmbH, Munich, Germany) measurements of hypofibrinogenaemia (fibrinogen 9/l) in 200 adult non-transplant patients during cardiopulmonary bypass (CPB). Blood samples were obtained for FIBTEM (assay for the fibrin part of the clot), PLTEM (calculated platelet-specific component), and laboratory measurements simultaneously. Our thresholds for FIBTEM and PLTEM were A10 (clot firmness 10 minutes after clotting time) ≤8 mm, and 8 mm and PLTEM A10 values ≥35 mm, but would continue to use laboratory measurements for confirmation. We would not use FIBTEM A10 ≤8 mm or PLTEM A10