Invasive management of atrial tachycardias(ATs) requires proper diagnosis of the mechanism followed by elimination of the responsible substrate. A novel lattice-tip catheter with both high-density mapping and dual ablation properties(radiofrequency-RF/pulsed field ablation-PFA) has been recently introduced for catheter ablation of atrial fibrillation. We present the first study to assess its performance in the management of ATs (diagnostic and therapeutic).
MethodsPatients with documented ATs were selected. Activation mapping was used for the establishment of the AT mechanism. Confirmation with entrainment was performed, whenever appropriate. Accuracy of the activation mapping in diagnosis, acute ablation efficacy, and procedural characteristics were the study endpoints.
ResultsTwenty patients were included (12 cavotricuspid isthmus-dependent atrial flutters, 5 mitral flutters, 2 roof flutters, and 2 focal ATs). Proper diagnosis was established by activation mapping in all cases. The mean mapping time was 7.85 ± 3.06 min with 296.82 ± 150.9 mean mapping points/minute. The mean ablation time was 54.25 ± 42.97 s. Conversion to sinus rhythm during ablation was achieved in all cases with the exception of a roof flutter that converted to mitral flutter and a case of a parahisian AT in which ablation was not attempted. Patients that received ablation did not experience any arrhythmia recurrence in a mean follow up of 4.14 ± 0.91 months. No major or minor complications occurred.
ConclusionThe lattice-tip catheter and its dedicated electroanatomical mapping system provided sufficiently detailed activation mapping for the diagnosis of the AT mechanism. The delivered lesions were highly effective acutely, with no adverse events. However, limitations exist and should be acknowledged.
Graphical Abstract
留言 (0)