Evaluation of cyclic fatigue behavior of RACE EVO rotary Files: In-vitro comparative study

The Saudi Dental Journal

Available online 18 March 2024

The Saudi Dental JournalAuthor links open overlay panel, , , , AbstractIntroduction

Comparative evaluation of the newly introduced; RACE EVO file system to the currently well-performing file systems like HyFlex EDM OneFile, Vortex Blue, ProTaper Gold (PTG) with regards to the cyclic fatigue resistance.

Materials

&Methods.

The endodontic rotary files under evaluation were segregated into four groups as follows:; RACE EVO, PTG, Vortex Blue, and HyFlex EDM OneFile. A detailed physical analysis of the cyclic fatigue resistance was performed in 15 files of each type (n = 60). Under the test conditions of 37 °C the test samples were subjected to continuous motion at the manufacturer recommended speed within an artificial curved canal until it fractured. For all the test samples, with the recording of time of fracture, calculation was done for number of cycles to fracture (NCF). In addition, the length of the fractured segment was determined. The fractured surface was further evaluated with scanning electron microscopic (SEM) images.

Results

Highest significance for NCF was seen in HyFlex EDM OneFile followed by the Vortex Blue, PTG, and; RACE EVO (p < 0.05) by analyzing the data statistically. Significant differences among the lengths of the fractured segments were evident among tested files, ranging between 4.73 and 6.3 mm (p < 0.05) with HyFlex EDM OneFile uniquely differentiating from the others showing the fracture at 6.3 mm. The common features of cyclic fatigue failure was further substantiated by SEM images.

Conclusions

Compared to the HyFlex EDM OneFile, Vortex Blue, and PTG files; RACE EVO demonstrated significantly lower resistance to cyclic fatigue.

Keywords

Cyclic fatigue

RACE EVO

HyFlex EDM

ProTaper Gold

Endodontic file

© 2024 THE AUTHORS. Published by Elsevier B.V. on behalf of King Saud University.