Do newborn infants exhale through the CPAP system? Secondary analysis of a randomised cross-over trial

WHAT IS ALREADY KNOWN ON THIS TOPICHOW THIS STUDY MIGHT AFFECT RESEARCH, PRACTICE OR POLICY

Knowledge on exhalation through the CPAP device is relevant for future studies comparing CPAP systems and interfaces. For clinicians, it could represent an opportunity for optimisation of CPAP treatment.

Introduction

Using non-invasive ventilation (NIV) as respiratory support is the standard of care in neonatal settings.1 2 Nasal continuous positive airway pressure (nCPAP) is the most widely used support for newborn infants with respiratory distress3 4 and has been shown to reduce morbidity and mortality in preterm infants.5 Early use of nCPAP is recommended by WHO as an important intervention to reduce neonatal mortality.6

There are multiple factors to consider when optimising nCPAP treatment. Clinical studies have focused on the optimal pressure during treatment, devices to use and different strategies when supporting neonates.7–10 Recent studies on nCPAP interfaces and failure of support, have favoured nasal mask over prongs.11 12

Historically, newborn infants were thought to be obligated nose breathers.13 This has later been rejected,14 15 and more recent trials have shown that infants can breathe through the mouth, making it more correct to describe them as preferred nose breathers.16

Even if infants prefer to breathe through the nose, we have not identified any studies that report infants actually exhaling through the nasal interface when treated with CPAP systems. The majority of nCPAP systems intended for neonatal use do not measure leakage or expiratory flow. In clinical settings, it is difficult to know if exhalation is through the nCPAP device or solely through leakage. Research on this topic could lead to improvements in nCPAP support providing details for improved management as well as interfaces and device design.

In the ToNIL cross-over trial, we measured absolute leakage for prongs and nasal masks in 50 infants on nCPAP (online supplemental file 1).17 We observed lower leakages with prongs and large variations in the degree of leakage between infants and between the two interfaces. The recorded flows from the ToNIL trial also allow us to measure expiratory flows and to determine if there is exhalation through the nCPAP system.

The aim of this study was to measure if infants exhale through the nCPAP system.

Materials and methods

This is a secondary analysis of data from our randomised crossover trial (ToNIL) that compared leakage during CPAP with two types of interfaces. The study population consisted of 50 infants recruited at two hospitals in Sweden, Östersund Hospital and Karolinska University Hospital in Stockholm. The inclusion criteria were nCPAP treatment with stable, spontaneous breathing and a postmenstrual age (PMA) >28 weeks. Exclusion criteria were respiratory or cardiac malformations, facial injuries, FiO2>0.5, recent surgery, circulatory instability or if the infant was recently extubated. Informed parental consent was obtained before inclusion.

The primary outcome was leakage measured in L/min and the site of leakage (through mouth or at the interface) was not determined. For each infant, the fresh gas flow and expiratory flow with both nasal prongs and nasal mask were measured using the flow-through technique,18 resulting in a total of 100 recordings. The flow-through technique was first described by Rigatto and Brady in 1972.18 It is used to determine absolute leakage and zeroing flow measurement devices. It has a high precision but the non-linearity and effects of humidity give a small error (<5%) which requires calibration.

After calibration with correct oxygen level, we added two custom-made SFM-3200-60-AW flow metres to the patient circuit, one on the fresh gas limb and another on the patient expiratory limb. The distal flow metres were zeroed against the fresh gas flow, while the interface connected was occluded. With this flow-through technique, there is no extra dead space added to the respiratory circuit.19

Details of the used methods and CONSORT diagram is available in online supplemental file 2 and in the original publication.17

Each measurement was 30 s long and was recorded while the infant was calm and breathing quietly. The nCPAP system was attached with prongs or nasal mask in a random order by an experienced nurse blinded to the measured leakage and pressure. During recording, the infant could be in an incubator, in a cot or skin-to-skin with either parent. All measurements were performed with devices of the original infant flow design (nFlow, Intersurgical, Berkshire, UK, or Inspire nCPAP, Inspiration Healthcare, Leicester, UK). All infants received humidified warm air during measurements. Respiratory curves were analysed retrospectively, using National Instruments LabVIEW 2015. Infant breathing and leakages were measured by subtracting the nCPAP device expiratory flow from the fresh gas flow. The system was calibrated and zeroed before each measurement. With no breathing and no leakage, the sum of flows was zero. With leakage, the flow curve was offset upwards. During breathing, the curve crossing zero (ie, a negative flow value) indicated that exhalation was through the CPAP system to some extent (figure 1). To determine if an infant was exhaling through the system, the respiratory recordings needed to show flow reversal in more than 50% of all breaths, each for a minimum duration of 0.1 s, during a 30 s recording. If all these criteria were met, the infant was considered as having exhaled through the CPAP system. We found no definition of exhalation based on measurement of flow and the selected cut-off values have not been used before.

Figure 1Figure 1Figure 1

Examples of recordings from an infant that exhaled through the CPAP with nasal mask but not with prongs. With prongs, the fresh gas flow was 6.61 L/min, delivered CPAP was 2.74 cmH20 and the recorded leakage was 5.4 L/min. No exhalation through the CPAP device was seen. With nasal mask, the fresh gas flow was 6.64 L/min, CPAP was 3.63 cmH20 and recorded leakage was 1.5 L/min. The infant expired through the CPAP system in all breaths. With low leakage (nasal mask), the pressure at the device during exhalation was higher than the set CPAP of 4 cmH2O (red dashed line). This infant was born after vaginal delivery, at a gestational age of 41+4 w+d and a birth weight of 3605 g. At the time of recording, the infant was 9 hours old, received nCPAP support for transient tachypnea of the newborn at 4 cmH20 with FiO2 0.21 and had a saturation 98%. The graphs display 10 s of the original 30 s recordings. The reported mean values are from 30 s. nCPAP, nasal continuous positive airway pressure.

Expiration through the CPAP system was not defined or registered as an outcome before starting the trial.

Statistical analyses were performed in SPSS V.27. Demographic data for study subjects were summarised using descriptive statistics. Outcome variables were reported as means (±SD) or median (IQR). Group comparisons were conducted using Mann-Whitney U tests, chosen because of skewed distributions in variables in both groups. A p value of <0.05 was considered statistically significant.

Results

All 50 infants were stable during measurements, with low oxygen demand. Background information is provided in table 1. There were no adverse events and measurements were well tolerated. The median time for study measurements and switching between interfaces for each patient was 25 min (IQR 21–30).

Table 1

Demographic data and characteristics of study participants

The majority, 32/50 neonates, met our criteria of exhalation through the nCPAP system in more than 50% of all breaths in at least one of the two recordings (nasal mask or prongs), and 18/50 did not. Since infants exhaling through the CPAP system with one interface did not necessarily exhale when measured with the other interface, this resulted in 43/100 recordings showing exhalation through the system (figure 2). In the remaining 57 recordings, exhalation through the CPAP was less than 50% of exhalations or did not reach a minimum length of 0.1 s. Leakages exceeding 0.3 L/min were seen in 97/100 measurements, with a maximum leakage of 9.9 L/min. In recordings where the infant exhaled through the nCPAP, leakages were lower (p<0.01, table 2).

Table 2

Comparison of exhalation through nCPAP for the two interfaces

Figure 2Figure 2Figure 2

Proportion of breaths with reversed flow for the 50 recordings with nasal mask and the 50 recordings with prongs. In 43 of 100 measurements, most of the recorded breaths revealed exhalations through the nasal continuous positive airway pressure system (bars to the right, >50% of breaths).

When comparing the 32 infants who exhaled through the nCPAP at any time with those who did not, infants who exhaled through the system had a shorter median duration of CPAP treatment and lower median postnatal age (table 3).

Table 3

Comparison of infants that exhaled through the CPAP system in either interface and those who did not

Analysing this for each interface showed no difference between groups when recorded with nasal mask, but for prongs recordings more exhalations were seen with lower postnatal age, higher weight and shorter duration of CPAP treatment (table 2).

Discussion

In this secondary analysis of infants in the ToNIL trial, we examined exhalation through the CPAP system. The main finding was that exhaling through the system is common and related to low leakage (table 3). Using flows to determine the proportion of infants exhaling through the CPAP system has not previously been reported.

The relation between exhaling through the system and leakage was expected. For infants not exhaling through the system, leakage was greater than the expiratory flow and exhalation was entirely through leakage. The cause of leakage could not be determined and could have been at the interface, through the mouth or to the oesophagus. The recordings were performed after application by a nurse, unaware of the delivered pressure or leakage. We have previously reported that leakage could be reduced with simple manoeuvres by an investigator guided by the measured leakage. Such guided manoeuvres could result in more infants exhaling through the system (online supplemental file 1).17

Comparison of infants that exhaled through the nCPAP and those who did not revealed no differences in GA or BW. There was a difference in postnatal age and CPAP duration between the groups (table 3), but when dividing the group based on interface this difference was significant for prongs but not for mask.

A common recommendation in CPAP care is to use well-fitting interfaces to minimise leakage and deliver an effective distending pressure in combination with attentive care to avoid nasal damage.20 Studies on leakage during CPAP treatment are few and no previous studies with blinded application of CPAP systems have been identified.

In a study from 2005, De Paoli et al 21 investigated mean pharyngeal pressure and mouth closure in 11 preterm infants treated with bubble CPAP (bCPAP). They report a reduction in mean pressures delivered to the pharynx compared to at the interface depending on if the mouth was closed or open. They observed that the mean pharyngeal pressures never exceeded the mean set CPAPs and there was always some pressure loss regardless of mouth position. They concluded that it was unlikely that infants exhale through the system and that nCPAP is unlikely to increase expiratory resistance. Their reported average loss of pressure corresponds to our finding that mean leakage was high and that few patients have no leakage. We believe that measuring flows gives a more accurate description when examining exhalation, compared with measuring pressures.

The clinical evidence for using well-fitting interfaces and minimising leakage is low but widely acknowledged as a part of delivering quality care for infants receiving CPAP. In our study, we used a conventional low resistance CPAP system with short binasal prongs or nasal mask. The infant flow device used is pressure stable and easy to exhale through. The function has been described as a fluidic-flip,22 with gas entrainment from a jet that can ‘flip’ between inspiration and expiration. Compared with other devices, this could have facilitated exhalation through the device.

When using interfaces that are not designed to provide a snug fit and have high expiratory resistance, it is important to bear in mind that these interfaces are not likely to allow exhalation through the CPAP system under any circumstance. An example of this is the RAM cannula which is sometimes used with CPAP devices off label.23 There are also examples of systems with short wide nasal prongs and nasal masks designed to be well fitting without leakage where the CPAP generator connected to the interface has high resistance.24 We believe that exhalation through these devices is less common but this was not examined in this study. If CPAP care and outcomes can be improved by minimising leakage and increasing exhalation through the CPAP system remains an open question.

Limitations

We used a CPAP system with low resistance and interfaces designed to fit without leakage.9 Other CPAP systems and interfaces might give different results. There was no available definition of what constitutes exhalation through a CPAP system. In our study, we arbitrarily defined exhalation through the system as the device expiratory flow exceeding the fresh gas flow for more than 0.1 s in more than 50% of expirations. The study was small and a major limitation was that the recordings were short. Even if the nurses were blinded to leakage measurements, the results reflect point prevalence in an experimental situation and recordings of hours or days during routine care would be of value.

Conclusion

During nCPAP treatment, infants treated with nCPAP can exhale through the CPAP system. For preclinical research, this is relevant for studies comparing CPAP systems and interfaces. For clinicians, it could represent an opportunity for optimisation of CPAP treatment by improving delivery of distending pressures. Using flows to determine exhalation through the CPAP system has not previously been reported.

Data availability statement

Data are available upon reasonable request.

Ethics statementsPatient consent for publicationEthics approval

This study involves human participants and was approved by the Swedish Ethical Review Authority, reference number 2016/2449-3. Participants gave informed consent to participate in the study before taking part.