Prospective Analysis of Versatility and User Satisfaction with a Novel Single-Use Cystoscope with Working Channel

Introduction: In the present study, a prospective systematic assessment of the clinical utility of the Ambu® aScopeTM 4 Cysto Reverse Deflection with regard to image quality, maneuverability, and navigation in an outpatient and inpatient setting was performed. Materials and Methods: A prospective multicenter study was performed for evaluation of the instrument during routine cystoscopy. We evaluated the clinical performance of the instruments using a standardized user questionnaire in different categories including image quality, treatment success, imaging of all areas of the urinary bladder, quality of navigation, flexibility of the endoscope, and satisfaction with the device. Statistical analyses were performed by SPSS using the Kruskal-Wallis and Wilcoxon-Mann-Whitney tests. A p value of p ≤ 0.05 was defined as statistically significant. Results: A total of 200 cystoscopies were performed, and the questionnaire response rate was 100%. The image quality was rated as very good in 65.5% (n = 131), good in 30.5% (n = 61), and neutral in 4% (n = 8) of cases. The criteria for poor or very poor were not mentioned. The characteristic “treatment success based on image quality” was also evaluated as very good in 49% (n = 98) and good in 50.5% (n = 101). The analysis revealed a very good or good overall impression of the examiners in all cases. Replacement of the cystoscope was not necessary during any of the examinations. However, in 3 cases, technical difficulties were documented. Further analysis of the data showed that physicians with less professional experience rated the visualization of the urinary bladder (p = 0.007) and the treatment success with regard to image quality significantly worse (p = 0.007). Conclusion: The Ambu® aScope™ 4 Cysto Reverse Deflection shows high satisfaction values among users in clinical routine. In analogy to other studies with flexible endoscopes, urologists with more professional experience show higher satisfaction values than examiners with less training in flexible endoscopy.

The Author(s). Published by S. Karger AG, Basel

Introduction

Cystoscopy is the most common endoscopic examination in urology. It is used for diagnostic purposes, e.g., for evaluation of the bladder mucosa in case of tumors of the bladder and urethra, as well as for interventional purposes, e.g., for forceps biopsies or for removal of inserted ureteral stents. Good endoscopic visualization and device handling are essential for the examiner.

The Ambu® aScope™ 4 Cysto Reverse Deflection is a new flexible disposable cystoscope certified for diagnostic cystoscopy in November 2020. Clinical studies show that the use of disposable endoscopes is more cost-effective compared to reusable endoscopes. Beebe et al. [1], Donato et al. [2], and Pietropaolo et al. [3] demonstrate a more favorable cost-effectiveness compared to conventional reusable cystoscopes [15]. The clinical advantages of single-use endoscopy also include the immediate availability of sterile disposable instruments. When using multiple endoscopes, hygienic aspects, a possibly limited number of available devices, and the time of reprocessing between two examinations must be taken into account in hospitals and outpatient offices [6].

Currently, however, there is no comprehensive validation with regard to the satisfaction of users and examiners with the handling and the device characteristics of the Ambu® aScope™ Cysto Reverse Deflection in clinical routine. Multicenter studies with other single-use devices primarily describe their clinical possible indications, e.g., based on the success rate of the removal of inserted DJ splints [7]. Despite poor flexion due to an instrument inserted in the working channel, Scotland et al. [8] described in an observational study a satisfying image quality using a single-use cystoscope. While systematic prospective-randomized trials comparing single-use cystoscopes versus reusable cystoscopes remain methodologically challenging, user satisfaction appears as a measurable effect size to determine the product quality and clinical applicability of a novel instrument. Therefore, in the present study, a systematic assessment of the clinical usability of the Ambu® aScope™ 4 Cysto Reverse Deflection integrating image quality, maneuverability, and navigation was performed in both an outpatient and inpatient setting.

MethodsStudy Population and Study Period

The evaluation was performed on a total collective of 200 patients who received a cystoscopy. Data collection took place between September 2021 and January 2022 at four centers (2 inpatient centers, 2 outpatient centers). A positive ethics vote for the observational study was granted by the Ethics Committee of the University of Tübingen under the number 387/2021B02.

Patients were included who had an indication for diagnostic cystoscopy, cystoscopic follow-up of previous urinary tract disease, biopsy of the urethral mucosa/bladder mucosa, botulinum toxin injection into the urinary bladder muscles, removal of ureteral stents, as well as for bladder stone removal in the course of a routine presentation in the clinic/practice. The Ambu® aScope™ 4 Cysto Reverse Deflection was used instead of a flexible reusable cystoscope for diagnosis/intervention. After the examination, the users were asked about image quality, navigability, and maneuverability using a standardized questionnaire. Patient-related clinical and histological variables were not included in the study.

Systematic Questionnaire

Baseline variables recorded were the investigators’ professional experience in two categories (1–10 and >10 professional years), the average number of cystoscopies performed by the investigators per week (1–2; 3–5, >5), the indication for performing cystoscopy, and any additional endoscopy accessories used. In addition, the following characteristics regarding examination quality were queried using a Likert scale (very good: 1, good: 2, neutral: 3, poor: 4, and very poor: 5): image quality, treatment success based on image quality, visualization of all areas of the urinary bladder, quality of navigation within the urinary bladder, quality of illumination, flexibility of the endoscope without endoscopy accessories, satisfaction with working channel size for biopsy, and flexibility of the endoscope with inserted endoscopy accessories. In addition, the image quality of the monitor, the menu navigation, data transfer options, and the 180° rotation possibility of the live image were evaluated. Finally, the question was asked about the overall impression regarding the image quality and functionality of the monitor as well as the overall impression regarding quality and functionality of the cystoscope (see Appendix 1: Questionnaire for users). If a change to another cystoscope was necessary during the procedure or if there were problems with the Ambu® aScope™ 4 Cysto, this was recorded.

The data were analyzed using SPSS Statistics version 28.0, and the scale levels were compared using the Kruskal-Wallis test and Wilcoxon-Mann-Whitney test. A significant difference was assumed at a value of ≤ 0.05.

ResultsBaseline Variables of the User Population

With 200 cystoscopies performed, the questionnaire response rate was 100%. Due to the anonymous questionnaire, the number of different investigators was not determined. In a total of 62.5% (n = 125) of responses, investigators reported professional experience of between 1 and 10 years, and 29.5% (n = 59) reported professional experience of more than 10 years. This information was missing from 16 questionnaires (8%). A total of 25.5% (n = 51) of respondents reported performing between 1 and 2 cystoscopies per week, 24.5% (n = 49) between 3 and 5 cystoscopies per week, and 44.5% (n = 89) more than 5 cystoscopies per week.

Indication for Cystoscopy

Table 1 summarizes the indication for cystoscopy during the observation period. Ambu® aScope™ 4 Cysto Reverse Deflection was used for the initial bladder examination in 47% (n = 94) of cases and for follow-up bladder examination in 38% (n = 76) of cases. Biopsies were performed in 3.5% (n = 7) of cases. Botox injections into the urinary bladder muscles were performed in 2.5% (n = 5) of cases. Urethral stents were removed in 4% (n = 8) of the examinations performed, and bladder stones were removed in 0.5% (n = 1) of the examinations.

Table 1.

Indication for cystoscopy

Frequency%Initial examination of the bladder9447.0Follow-up examination of the bladder7638.0Biopsy of the bladder and/or bladder neck73.5Botox injection52.5Removal of ureteral stent84.0Removal of bladder stone10.5Other interventions94.5Total200100.0Results of the User Survey

A detailed overview of the results of the user survey is shown in Table 2, and a summary is given in Table 3. The image quality was rated as “very good” in 65.5% (n = 131), “good” in 30.5% (n = 61), and “neutral” in 4% (n = 8) of cases. The criteria for poor or very poor were not mentioned. The characteristic “treatment success based on image quality” was also described as “very good” in 49% (n = 98) and “good” in 50.5% (n = 101). The visualization of all areas of the urinary bladder was “very good” in 48.5% (n = 97) of cases. The bending angle of the Ambu® aScope™ 4 Cysto was described as “very good” in 123 cases (42.8%) and “good” in 68 (34.7%) of cases. With inserted endoscopy accessories, the value “very good” was achieved in 16.9% and the assessment “good” in 38 cases (64.4%). In three cases of ureteral stent removal, the categories “poor” or “very poor” were selected. The working channel size for performing a biopsy was described as “very good” in 8 cases (33.3%), “good” in 12 cases (50%), and “neutral” in 3 cases (12.5%) or poor in 4.2% of cases. Switching to another cystoscope was not necessary in any examination. In the documentation of three cases with technical difficulties, it was observed that the irrigation flow could not be controlled with inserted accessories. A second three-way stopcock became necessary for this purpose. In addition, insufficient angulation of the instrument occurred during the removal of an inserted DJ splint with inserted endoscopy accessories. Another user described an insufficient inversion for evaluation of the bladder neck. The overall impression and functionality of the monitor and the single-use cystoscope were “very good” or “good” in all cases. A comparative analysis dividing the collective according to the professional years of the users showed that the visualization of all areas of the urinary bladder and the treatment success due to the image quality were rated lower in users with less professional experience (Table 4). The stratification according to the number of endoscopies performed per week shows a similar picture here: satisfaction with the Ambu® aScope™ 4 Cysto was rated better among users with a higher number of endoscopies per week (Table 5).

Table 2.

Results from the user survey

Very goodGoodNeutralBadVery badNot specifiedImage quality13165.5%6130.5%84.0%00.0%00.0%00.0%Treatment success due to image quality9849.2%10150.8%00.0%00.0%00.0%00.0%Imaging of all areas of the bladder9749.0%9347.0%63.0%21.0%00.0%00.0%Navigation inside the bladder12763.5%6130.5%126.0%00.0%00.0%00.0%Lighting13567.8%5628.1%84.0%00.0%00.0%00.0%Flexibility without endoscopy accessories12362.8%6834.7%52.6%00.0%00.0%00.0%Working channel size for biopsy833.3%1250.0%312.5%14.2%00.0%00.0%Flexibility with endoscopy accessories1016.9%3864.4%813.6%23.4%11.7%00.0%Image quality monitor13668.0%4824.0%157.5%10.5%00.0%00.0%Menu navigation8241.2%9849.2%199.5%00.0%00.0%00.0%Data transfer option5361.6%2731.4%55.8%11.2%00.0%00.0%180° rotation option of the live image3153.4%2237.9%35.2%00.0%00.0%11.7%Total impression quality + functionality Ambu aView 2 Advance13367.2%6432.3%10.5%00.0%00.0%00.0%Total impression quality + functionality Ambu aScope 4 Cysto13266%6834%00.0%00.0%00.0%00.0%Table 3.

Summary of results from the user survey

NMinimumMaximumAverage valueStandard deviationImage quality200131.390.564Treatment success due to image quality199121.510.501Imaging of all areas of the bladder198141.560.608Navigation within the bladder200131.420.605Lighting199131.360.559Flexibility without endoscopy accessories196131.400.540Working channel size for biopsy24141.870.797Flexibility with endoscopy accessories59152.080.772Menu navigation199131.680.640Data transfer option86141.470.663180° rotation option of the live image57031.470.630Total impression quality + functionality Ambu aView 2 Advance198131.330.483Total impression quality + functionality Ambu aScope 4 Cysto200121.340.475Table 4.

Comparison of professional years

Professional yearsNAverage rankpImage quality1–1012589.170.133>105999.55Total184Treatment success due to image quality*1–1012598.200.007*>105878.63Total183Imaging of all areas of the bladder*1–1012497.810.007*>105878.00Total182Navigation within the bladder1–1012592.060.847>105993.43Total184Lighting1–1012590.210.408>105895.86Total183Flexibility without endoscopy accessories1–1012387.300.150>105797.41Total180Working channel size for biopsy1–10914.170.328>101511.50Total24Flexibility without endoscopy accessories1–104231.460.227>101726.38Total59Image quality monitor1–1012588.940.104>1059100.03Total184Total impression quality + functionality Ambu aView 2 Advance1–1012389.960.481>105994.71Total182Total impression quality + functionality Ambu aScope 4 Cysto1–1012589.970.250>105997.86Total184Table 5.

Comparison cystoscopies/week

Number of cystoscopies per weekNAverage rankpImage quality*a. 1–2/week51110.39(c)b. 3–5/week49122.33(c)c. >5/week8971.13(a) (b)Total189<0.001*Treatment success due to image qualitya. 1–2/week5187.050.316b. 3–5/week48101.33c. >5/week8995.08Total188Imaging of all areas of the bladder*a. 1–2/week5078.48(a)b. 3–5/week48109.17(b)c. >5/week8994.54Total1870.006*Navigation within the bladder*a. 1–2/week5192.26(b)b. 3–5/week49117.56(a) (c)c. >5/week8984.15(b)Total189<0.001*Lighting*a. 1–2/week51108.03(c)b. 3–5/week48121.37(c)c. >5/week8972.25(a) (b)Total188<0.001*Flexibility without endoscopy accessories*a. 1–2/week47117.66(c)b. 3–5/week49120.64(c)c. >5/week8967.90(a) (b)Total185<0.001*Working channel size for biopsya. 1–2/week1114.000.070b. 3–5/week1013.25c. >5/week34.50Total24Flexibility with endoscopy accessoriesa. 1–2/week1431.680.282b. 3–5/week1334.65c. >5/week3227.38Total59Image quality monitor*a. 1–2/week51110.59(c)b. 3–5/week49118.83(c)c. >5/week8972.95(a) (b)Total189<0.001*Total impression quality + functionality Ambu aView 2 Advance*a. 1–2/week50111.10(c)b. 3–5/week48122.13(c)c. >5/week8969.22(a) (b)Total187<0.001*Total impression quality + functionality Ambu aScope 4 Cysto*a. 1–2/week51113.53(c)b. 3–5/week49121.36(c)c. >5/week8969.87(a) (b)Total189<0.001*Discussion

With the introduction of single-use endoscopes in urology, alternative systems to reusable cystoscopes are available, which are currently under scientific validation in terms of quality, cost-effectiveness, and ecology. Available studies with single-use cystoscopes (e.g., for the removal of ureteral stents) have demonstrated their clincal efficacy and cost-effectiveness compared with reusable cystoscopes [13]. The aim of our study was to investigate the satisfaction of users and examiners with the handling and features of the Ambu® aScope™ 4 Cysto Reverse Deflection in everyday clinical practice. The evaluation of the results showed a very good or good overall impression of the examiners in 100% of the cases. In none of the examinations performed was it necessary to switch to an alternative system in order to complete the examination successfully. In 2022, Whelan et al. [9] showed that there was no technical or optical disadvantage between the Ambu® aScope™ 4 Cysto Reverse Deflection and the sterilizable Olympus SD scope. With the exception of the 3 mm distance, where an optical advantage of the compared digital reusable cystoscope was observed. Scotland et al. [8] studied a single-use cystoscope (NeoScope) and found that flexion, image quality, and maneuverability were equivalent to the reusable cystoscopes. Disadvantageously, reduced flexion with instruments in the working channel was described for single-use cystoscopes.

In the present study, technical difficulties were noted in 3 of the 200 examinations performed. E.g., inversion was not sufficiently possible to assess the bladder neck in one case. In this case, the examination was performed by an examiner with low professional experience (1–2 cystoscopies per week, professional experience less than 10 years).

Schout et al. [10] evaluated beginners and advanced examiners with an endoscopy simulator and found that with increasing experience, the quality of flexible cystoscopy improves and the time required to perform a qualitative examination decreases. In order to perform a sufficient examination, a threshold of 122 cystoscopies to full quality in flexible cystoscopy was defined in another study [11]. The described difficulties in handling the equipment could therefore also be due to the low professional experience of the examiner. In the same context, examiners in the present study with higher professional experience gave a significantly better overall evaluation of the disposable cystoscope than examiners with less experience.

Furthermore, a lack of flexion of the instrument with inserted grasping forceps for the removal of an inserted DJ was documented. In this case, the examiner cited that the available grasping forceps for flexible cystoscopy were too short for the disposable cystoscope. Despite the fundamentally high compatibility of single-use instruments with endoscopic aids, sufficient passage of the working channel forceps and baskets should be checked as part of the acquisition. Here, the working channel of the Ambu aScope 4 Cysto has a diameter of 6.6 Fr. and is thus comparable to available flexible urethrocystoscopes.

In an earlier study, Wong et al. [12] focused their analysis on patient satisfaction with the Ambu® aScope™ 4 Cysto Reverse Deflection. All patients had previous experience with a reusable cystoscope. The authors described that 95% of the subjects would prefer the single-use cystoscope over the reuseable cystoscope due to the shorter procedure time.

Apart from the general aspects of using single-use cystoscopes or reusable endoscopes like logistic aspects and immediate availability of endoscopes [15], costs and cost-effectiveness of the procedures [13, 14], and hygienic aspects [1517], increasing attention has recently been given to waste and environmental issues related to sterilizable systems and single-use instruments as part of sustainability efforts and “planetary health” in healthcare. Su et al. [4] reported costs for using reusable cystoscopes to remove DJ splints in a range from $165 to $1,469 in dependence of the number of investigations performed. In comparison, the cost of single-use cystoscopy was reported between $244 and $420. The authors concluded that for facilities performing fewer than 700 cystoscopies annually with a total of 20 multiple cystoscopes available, greater cost-effectiveness can be achieved using single-use cystoscopes. Davies et al. [18] investigated the carbon footprints of single-use URS and reusable systems and surprisingly found no significant differences between the devices (4.43 vs. 4.47 kg CO2 per device/use). However, the type of re-sterilization and the associated energy consumption for the reusable devices appear most relevant here. By contrast, the disposal of single-use devices is primarily responsible for the respective CO2 emissions. It should be kept in mind that apart from reprocessing and disposal of the medical products, transportation and logistic efforts represent highly relevant parameters for CO2 emissions in urological endoscopy.

Despite its prospective design, the present study is not without limitations. A main limitation of the present approach to user quality assessment is the lack of a control group, so that no comparative results with reusable endoscopes could be obtained. Noteworthy, a comparative study of reusable and single-use instruments is challenging in the context of a multicenter study given a diverse distribution of reusable endoscopes from different manufacturers, different imaging technologies (chip on tip or classic), and investigator experience in the respective centers.

While the analysis of single-use versus reusable endoscopy in urology is an open scientific field with ongoing research and discussion, the clinical data here demonstrate the clinical feasibility of single-use endoscopy using the Ambu® aScope™ 4 Cysto Reverse Deflection based on an extensive user survey.

Conclusion

The Ambu® aScope™ 4 Cysto Reverse Deflection shows high satisfaction values among users in clinical routine. In analogy to examinations with other flexible devices, examiners with more professional experience show higher satisfaction values than colleagues with less experience. A change to an alternative system was not necessary during the study.

Acknowledgments

We wou