Low-dose fluoroscopy technique drastically decreases patient radiation exposure during percutaneous nephrolithotomy

Fluoroscopy is essential in percutaneous nephrolithotomy (PCNL) but exposes patients and operating room staff to radiation. We investigated whether a low-dose (LD) protocol could reduce radiation exposure during fluoroscopy-guided access without compromising clinical outcomes. Patients undergoing PCNL with fluoroscopy-guided access at a tertiary care stone center between January 2019 and July 2021 were identified. Prior to September 3, 2020, the Philips Veradius C-arm’s default settings were used: standard per-frame dose, 15 pulses per second (PPS) frame rate. After this date, a low-dose protocol was used: reduced per-frame dose, reduced frame rate of 8 PPS for needle puncture and 4 PPS for all other steps. Clinical and radiographical data were retrospectively collected. The primary outcome was cumulative radiation dose. Secondary outcomes were stone-free status (SFS; defined as no fragments ≥ 2 mm) and complications. Multivariate regression analysis was performed. 100 patients were identified; 31 were in the LD group. The LD cohort was exposed to a significantly lower mean cumulative radiation dose of 11.68 mGy compared to 48.88 mGy (p < 0.0001). There were no differences in operative time, fluoroscopy time, stone burden, SFS, or complications. In a multivariable regression model adjusting for several variables, LD protocol was associated with lower radiation dose while skin-to-calyx-distance (STCD) was positively associated with cumulative radiation dose. Low-dose fluoroscopy and decreased frame rate during PCNL decreased radiation exposure fourfold without affecting SFS or complication rates.