Tomography, Vol. 8, Pages 2864-2873: Evaluation of Anterior and Posterior Corneal Higher Order Aberrations for the Detection of Keratoconus and Suspect Keratoconus

1. IntroductionKeratoconus (KC) is a progressive corneal disease characterized by stromal thinning and protrusion, resulting in irregular astigmatism and visual impairment [1]. Ectasia may initially be present unilaterally, however due to the progressive nature of the disease, both eyes are often affected [2]. Advanced signs of KC can be detected clinically using slit-lamp biomicroscopy as well as keratometry. Visual impairment in such cases is common. Screening can lead to early detection of the disease and ectasia. Split or scissoring retinscopic reflex, or the “Charleaux” oil droplet sign are clinical indicators that may be seen in early keratoconus [3].The use of videokeratography-derived indices is the most sensitive and sophisticated way of detecting KC. However, some corneal anomalies induced by corneal scarring, dry eye, or hard contact lenses, may yield false positives on videokeratoscopy and findings ordinarily seen in early keratoconus. Similarly, a negative videokeratography may not rule out the absence of an early form of KC [4]. Several studies have suggested that the addition of corneal aberrometry data may allow for better sensitivity and specificity for the detection of keratoconus in comparison to topographical evaluation alone [5,6,7].

The aim of this study was to determine the diagnostic accuracy of the anterior and posterior corneal wavefront aberrations in differentiating between keratoconus, suspect keratoconus (SKC), and normal corneas, and to ascertain the parameters with the highest sensitivity and specificity.

2. Material and Methods

This single-centre, retrospective, case-controlled study was conducted at the department of ophthalmology, Tishreen University Hospital (Latakia, Syria). Study participants underwent comprehensive ocular assessment and included: uncorrected distance visual acuity (UDVA), corrected distance visual acuity (CDVA), auto-refracto-keratometry (SEIKO CO., GR-3500KA, Tokyo, Japan), slit-lamp examination to detect the presence or absence of signs of KC (apical scar or thinning, Fleischer rings and Vogt’s striae), Goldmann applanation tonometry, dilated funduscopy, and retinoscopic examination.

Corneal tomography and aberrometry were conducted using the Sirius Sheimpflug-Placido topographer (Costruzioni Strumenti Oftalmici, Florence, Italy; software version: Phoenix v.2.6). Anterior corneal higher order aberrations (HOAs) were measured over the central 6.0 mm zone. Three well-focused, centered images were obtained for both eyes. The tear film was optimized in dry eye patients with topical lubricants or spontaneous repeated blinking. A quality threshold of 16 continuous Placido disc mires was set, to maintain image acquisition quality and subsequent calculation of the Zernike coefficients for a 6.0 mm simulated pupil. All subjects with a history of contact lens use were asked to discontinue wearing their lenses for 3 weeks and 1 week for rigid lenses and soft lenses, respectively.

2.1. Study Participants

The 228 eyes of 188 patients were included and divided into three study groups: normal, SKC, and KC.

The normal group was composed of individuals who had undergone previous laser-assisted in situ keratomileusis (LASIK) at least 3 years prior with no evidence of postoperative ectasia. Only the preoperative data were considered in the normal group. All eyes in the normal group had a corrected distance Snellen (feet) visual acuity of 20/20 or better. Only data from one eye (right eye) was included in the normal group. Suspect KC was defined as anterior or posterior corneal steepening, absence of clinical (keratometric, retinscopic, or biomicroscopic) signs of keratoconus in either eye, and best corrected visual acuity of 20/20 or better [8,9]. A diagnosis of KC was made if (a) there was an irregular cornea, determined by distorted keratometry mires or distortion of the dilated retinoscopic reflex (or a combination of both) [10,11], in addition to (b) at least two of the following tomographic/topographic findings: abnormal posterior ectasia determined by symmetry index back (SIb) of >0.46 D, abnormal thickness distribution, or symmetry index front (SIf) of >1.17 D [4]; or one of the following slit lamp findings: Vogt striae, 2-mm arc of Fleisher ring, or corneal scarring consistent with KC [12].

Exclusion criteria included patients with a history of previous corneal or ocular surgery (e.g., cataract, glaucoma, corneal cross-linking, excimer laser surgery, intrastromal corneal rings, phakic intraocular lens), ocular co-morbidities that could interfere with the readings/results (e.g., dry eye disease, keratitis, glaucoma, uveitis, corneal endothelial dystrophy), corneal scarring not consistent with KC, autoimmune disease, lactation, or pregnancy. Individuals with ocular or systemic disease (e.g., uveitis, glaucoma, atopic dermatitis, connective tissue disorders) were also excluded from the normal group.

2.2. Mean Outcomes MeasuresAnterior and posterior corneal HOAs were obtained from the Sirius tomographer at a 6 mm diameter. Normalized coefficients were used, expressed in microns of wavefront error (root mean square [RMS]), and labeled with International Organization for Standardization (ISO) standardized double-index Zernike (Z) symbols [13]. The collected data included: RMS total HOAs, RMS trefoil Z (3, ±3), RMS coma Z (3, ±1), and RMS primary spherical aberrations (SA) Z (4, 0). 2.3. Statistical Analysis

The study was approved by the research committee of Tishreen University (approval number TUH-23022) in accordance with the ethical standards stated in the 1964 Declaration of Helsinki, with the informed consent of participants. Statistical analysis was performed using MedCalc Statistical Software (version 19.5.3, MedCalc Software, Ostend, Belgium) and SPSS software (version, 17, SPSS Inc, Chicago, IL, USA). The ANOVA test and Student’s t-test were used to compare variables of three and two groups, respectively. In cases when the data was not normally distributed, the Kruskal–Wallis, and Mann–Whitney tests were used to compare three and two groups, respectively. Receiver operating characteristic (ROC) curves were used to distinguish KC and SKC from normal corneas. These curves were obtained by plotting sensitivity against 1-specificity, calculated for each value observed. The area under the ROC curve (AUC) measures discrimination, which is the ability of the test to accurately classify eyes with and without disease. An area of 1.0 represents a perfect test, whereas an area of 0.5 represents a poor test. A p-value of 0.05 or below was considered statistically significant.

4. DiscussionAmongst different factors proposed to predict the risk of developing post-LASIK ectasia, early keratoconus was considered the main risk factor. [14]. Detecting corneas with suspect keratoconus is crucial in preventing post-LASIK ectasia. Initially, the term keratoconus suspect was introduced to describe eyes that had demonstrable evidence of subtle Placido-based videokeratography abnormalities without clinical evidence of disease [9].While Placido-based topography analyses the central anterior corneal surface, Scheimpflug-based imaging provides data from the anterior and posterior cornea and produces a full corneal thickness map. Several studies have shown that corneal HOAs are higher in keratoconus eyes in comparison to normal eyes. Therefore, the combination of wavefront analysis and videokeratography may help to define keratoconus and increase the sensitivity and specificity for early detection of suspect keratoconus [7,15].

The aim of this study was to determine the diagnostic ability of anterior and posterior corneal HOAs measured by the Sirius Scheimpflug tomographer in differentiating suspect keratoconus, keratoconus, and normal corneas. It also aimed to ascertain the parameters with the highest sensitivity and specificity.

We evaluated corneal HOAs that are most relevant to clinical practice including coma, trefoil, and spherical aberrations. Maeda et al. compared corneal aberrations in normal eyes and eyes with forme fruste or mild keratoconus, and found that coma-like and spherical-like aberrations were significantly higher in the latter group [16]. Our results showed a significant difference between the control and suspect keratoconus group, as well as between the control and keratoconus group for the anterior corneal RMS HOAs (total, trefoil, coma-like, and spherical HOAs). When comparing our results to the results of Alió and Shabayek who evaluated the anterior corneal RMS HOAs in normal and keratoconic eyes [17], the results confirmed significantly higher anterior corneal aberrations in keratoconic eyes compared to normal eyes. In our study, the mean anterior RMS HOAs values in the keratoconic group were lower (2.11 ± 0.9 µm) than those of Alió and Shabayek (3.14 ± 1.64 µm). This can be explained by the fact that the patients evaluated by Alió and Shabayek had more advanced keratoconus than those evaluated in our study.The increase in coma-like aberrations in keratoconic eyes is related to the cone decentration resulting in decentration of the visual axis of the eye. Jafri et al. found that RMS coma was 0.229 µm for normal eyes, 0.639 µm for eyes with suspected keratoconus, and 2.034 µm for eyes with early keratoconus [7]. Our findings are consistent with this study, despite different acquisition methods, namely the Sirius tomographer used in our group and the Alcon LADARWave used by Jafri et al. Moreover, our results agree with Alió and Shabayek, both confirming the increase in the total HOAs in the keratoconic group was primarily due to coma-like aberrations.In their study, to develop a keratoconus detection scheme based on Zernike coefficients, Gobbe and Guillon found that the best anterior HOAs detector at differentiating between suspected keratoconus and normal corneas was vertical coma (C3, −1) with a specificity of 71.9% and sensitivity of 89.3% [18]. Saad and Gatinel found that the best cut-off value for anterior corneal RMS coma to differentiate forme fruste keratoconus from normal corneas was 0.157 µm with a sensitivity of 71% and a specificity of 78% [15]. In contrast, our results showed that RMS coma with a cut-off value of >0.2 µm was the parameter with the highest AUC (0.992) to distinguish between suspect keratoconus and normal eyes with a sensitivity of 95.24% and a specificity of 75%. The difference in the sensitivity of discrimination of corneal coma between the two studies is due to the fact that our study group objectively selected suspect keratoconus, while Saad and Gatinel selected forme fruste keratoconus eyes (normal fellow eyes of keratoconus patients). Forme fruste keratoconus is the earliest identifiable stage of keratoconus.Our results showed that all anterior corneal HOAs indices except spherical aberrations had a good ability to detect keratoconus with AUC > 0.9. However, the greatest ability was seen for total HOAs and coma (AUC = 1 for both) with cut-off points of >0.65 and >0.57, respectively. Buhren et al. reported that corneal vertical coma (Z3, −1) can be used to distinguish keratoconus from normal eyes (AUC = 0.980; cut-off, −2.00 μm) [5]. However, the cut-off value of these aberrations was not similar to our study. The cut-off value depends on the stage of the disease and on the method of calculation of the aberrations. In the present study, we used the RMS value which represents the sum of horizontal and vertical coma aberrations, regardless of the orientation.In agreement with the theoretical optical properties of the corneal surface, we found that the posterior corneal surface was less aberrated than the anterior corneal surface. The refractive indices between air (1.0) and the anterior corneal surface (1.376), and between the aqueous (1.336) and the posterior surface (1.376) are not similar. This converts to about 1/14 of refraction occurring posteriorly compared to anteriorly [19]. Naderan et al. [20] and Xu et al. [21] indicated the importance of posterior corneal surface aberrations to differentiate normal from suspect keratoconus corneas. They found that the values for posterior coma of the normal group were 0.032 ± 0.363 and for the suspect keratoconus group were 0.193 ± 0.264 with statistically significant differences between groups. Bühren et al. found that vertical coma (C3, −1) was the only posterior surface parameter with a significant difference between suspect keratoconus and normal eyes [5]. However, they found no single posterior-surface coefficient nor RMS value that correctly classified ≥80% of the suspect keratoconus eyes and normal eyes. In the present study, we did not observe significant differences in the RMS of posterior total HOAs, trefoil, and coma-like HOAs between the normal group and the suspect keratoconus group. However, spherical aberration was the only parameter with a significant difference between suspect keratoconus and normal eyes (p 0.9 for all except RMS spherical aberrations, AUC = 0.846).

The present study has some limitations. The retrospective nature of the study is designed to analyse pre-existing data and is subject to bias. In addition, the lack of total ocular aberration data as a wavefront device was not used in this study. Future studies that combine total wavefront data, corneal higher order aberrations data, and tomography data may provide a better approach for the detection of corneas with the earliest stage of keratoconus and to predict corneas susceptible to ectasia.

In conclusion, although anterior and posterior corneal higher order aberrations data measured by the Sirius analyzer are very effective in distinguishing between keratoconus and normal eyes, only coma and total HOAs derived from the anterior corneal surface can be used to detect suspect keratoconus.