Successful trans-femoral retrieval of a stuck axillary Impella 5.0 device

We describe a case in which an axillary Impella 5.0, stuck in an area of calcification in the right subclavian artery, could not be retrieved in the usual manner. However, it was successfully removed using a long 22-Fr sheath and snaring catheter by means of the trans-femoral artery. Device retrieval using the trans-femoral artery snare technique is considered a valid option for removing the Impella device in patients who exhibit this complication.