Successful reintervention using dual-channel endoscope for perforation by partially migrated stent after endoscopic ultrasound-guided hepaticogastrostomy

  SFX Search  Permissions and Reprints

Endoscopic ultrasound-guided hepaticogastrostomy (EUS-HGS) is the standard endoscopic procedure for biliary obstructions [1] [2]. Although various complications such as migration, perforation, and abdominal peritonitis have been reported after EUS-HGS [3] [4] [5], the reintervention method for them has not been established. Herein, we present a case of successful reintervention for a migrated stent after EUS-HGS using a dual-channel endoscope and grasping forceps.

An 88-year-old man with pancreatic cancer underwent EUS-HGS for malignant biliary obstruction, and a partially covered self-expandable metallic stent (SEMS) was placed on the B3 branch. The patient had a high fever 2 days after EUS-HGS. Computed tomography (CT) revealed free air in the abdominal cavity ([Fig. 1]), which was caused by the shifting of the uncovered part of the SEMS to the gastric side by respiratory fluctuation. A reintervention for additional stent placement was urgently performed. A dual-channel endoscope (GIF-2TQ260M; Olympus, Tokyo, Japan) was inserted into the EUS-HGS site. A guidewire was then advanced through the EUS-HGS stent into the right hepatic duct ([Video 1]). However, the catheter could not advance through the B3 branch due to its strong bend. The EUS-HGS stent was too long, poorly anchored, and was difficult to align with the axis; therefore, it could not transmit force in the direction of the catheter ([Fig. 2]). Grasping forceps inserted from the other channel grasped and pulled the stent ([Fig. 3]), after which the catheter and stent delivery system advanced through the bend of the B3 branch by counter-traction ([Fig. 4]). The fully covered SEMS (Boston Scientific, Marlborough, Massachusetts, USA) was then deployed through the stent to the proximal B3 branch ([Fig. 5]). After 1 week, CT showed the disappearance of free air and the biliary metallic stent in the appropriate position. Stent placement using a dual-channel endoscope and grasping forceps may be considered a useful treatment for stent complications after EUS-HGS.

Fig. 1 Computed tomography images. a No free air in the abdominal cavity immediately after endoscopic ultrasound-guided hepaticojejunostomy (EUS-HGS). b The presence of free air 2 days after EUS-HGS.

Video 1 As a reintervention for additional stent placement, a dual-channel endoscope and grasping forceps were used. Pulling the endoscopic ultrasound-guided hepaticogastrostomy stent by grasping the forceps, catheter, and stent delivery system enabled passage through the bend of the B3 branch by counter-traction.

Fig. 2 Catheter could not advance through the strong bend of the B3 branch. a Fluoroscopic view. b Schema. Fig. 3 Endoscopic view showing the stent grasped using grasping forceps inserted from the other forceps channel of the dual-channel endoscope. Fig. 4 Fluoroscopic view showing the passage of the catheter by counter-traction through the strong bend of the B3 branch by pulling the grasped stent. Fig. 5 a Fluoroscopic view showing the deployment of the additional biliary metallic stent beyond the bend of the B3 branch to the proximal site in a stent-in-stent manner. b Endoscopic view showing the additional biliary metallic stent.

Endoscopy_UCTN_Code_CPL_1AL_2AD

Endoscopy E-Videos is an open access online section, reporting on interesting cases and new techniques in gastroenterological endoscopy. All papers include a high quality video and all contributions are freely accessible online. Processing charges apply (currently EUR 375), discounts and wavers acc. to HINARI are available.

This section has its own submission website at https://mc.manuscriptcentral.com/e-videos

Publication History

Article published online:
25 November 2022

© 2022. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/)

Georg Thieme Verlag KG
Rüdigerstraße 14, 70469 Stuttgart, Germany

留言 (0)

沒有登入
gif