Predictive Value of Transimpedance Matrix Measurements to Detect Electrode Tip Foldover

Objective 

To evaluate the ability of the transimpedance matrix (TIM) measurement to detect cochlear implant electrode tip foldover by comparing results to a “gold standard,” the intraoperative plain film radiograph.

Study Design 

Retrospective case series.

Setting 

Tertiary referral hospital.

Patients 

One hundred three patients who underwent cochlear implantation between June 2020 and August 2021.

Interventions 

Intraoperative electrophysiologic monitoring (electrode impedances, neural response telemetry, and TIM measurement) and modified Stenver's view plain film radiographs.

Main Outcome Measures 

Identification of tip foldover on both TIM and plain films.

Results 

In total, 103 patients (117 ears) had both a TIM measurement and intraoperative X-ray available for review, including 68 adults and 35 children. One hundred patients (85%) received the Cochlear Slim Modiolar electrode. Tip foldovers were noted in three of 117 implants (2.5%). In all cases, TIM was able to detect the foldover, and the electrode arrays were reinserted with the patients still under anesthesia, with repeat X-ray demonstrating a normal configuration. Two other abnormal TIM patterns were identified. One was in a patient with an obstructed cochlea in whom only 10 electrodes could be inserted, the other was in a patient with a common cavity abnormality. One additional patient underwent electrode repositioning intraoperatively because of overinsertion. In this patient, the TIM appeared to be within normal limits, but the over-insertion was apparent on X-ray. Overall, the sensitivity and specificity of TIM measurements in detecting electrode tip foldover were both 100%.

Conclusion 

TIM measurements were able to accurately identify tip foldovers. More research is needed to define the adjunctive role of TIM as an intraoperative measure.