Development and validation of a high-performance thin-layer chromatography densitometric method for the simultaneous determination of novel 1-acridinyl-1,2,3-triazole derivatives

Synthesis of targeted 1-acridinyl-1,2,3-triazole derivatives as anticancer agents

The targeted 1-acridinyl-1,2,3-triazole derivatives 6–10 were synthesized via a 1,3-dipolar cycloaddition reaction of 9-azidoacridine with the propargyl small molecules (R = 15) under click reaction conditions [7] as indicated in Scheme 1.

Biological evaluationIn vitro cytotoxicity activity of the target compounds

The hybrid compounds 6–10 were screened by the National Cancer Institute (NCI) (Bethesda, MD, USA) in a primary anticancer assay in the full NCI-60 cell panel. The cell lines were derived from nine different cancer types: melanoma, leukemia, colon, lung, central nervous system, renal, ovarian, breast, and prostate cancers. Compounds 9 and 10 having naproxen and valproic acid fragments showed low cell growth inhibition against all cancer cell lines. However, compounds 6–8 having 6-methyluracil, uracil, and phenytoin fragments displayed cell growth inhibition and selectivity against the different cell lines. Compound 6 displayed good to excellent cell growth inhibition against five types of cancer cell lines: leukemia (K-562, HL-60 [TB], and SR), melanoma (LOX IMVI), ovarian (IGROV1), prostate (DU-145), and breast (MCF7) with percent growth inhibition of 83.7%, 92.9%, 88.6%, 84.8%, 67.0%, 84.3%, and 97.5%, respectively. For compound 7, cell growth inhibition ranging from 61.3 to 67.0% was observed against K-562 and SR, MCF-7 and T-47D cell lines. Compound 8 showed comparable cell growth inhibition percent ranging from 60.8 to 68.6% against K-562, SR, and MDA-MB-468; however, it showed marked selectivity and the best inhibitory activity of 84.1% against NCI H522 (non-small cell lung cancer) [7].

MTT assay for cell viability

The cytotoxicity of compounds 6–8 against MCF-7 breast cancer and DU-145 prostate cancer cell lines was determined using the MTT assay. The compounds were all more cytotoxic in the MCF-7 cell line and in both cell lines the 6-methyuracil derivative 6 was more cytotoxic than the uracil derivative 7 with the bulk phenytoin derivative 8 the least active (Table 2). Compound 6 displayed comparable cytotoxicity (IC50 2.70 ± 0.08 mM) with the standard doxorubicin (IC50 2.06 ± 0.04 mM), an anthracycline in clinical use as a wide spectrum antitumor agent, which exerts its action by intercalation between DNA double strands and blocks the topoisomerase IIB involved in the DNA replication process [7].

Table 2 Cytotoxicity profile and topoisomerase IIB inhibitory activity of compounds 6‒8Topoisomerase IIB assay

The acridine ligands 6–8 were evaluated for inhibitory activity against MCF7 topoisomerase IIB, and compounds 6 and 7 showed potent inhibitory activity with IC50 of 0.52 mM and 0.86 mM, respectively, compared to doxorubicin IC50 0.83 mM, while the phenytoin derivative 8 was less active with IC50 2.56 mM (Table 2). Cell-cycle analysis of MCF7 and DU-145 cells treated with the active ligands 6–8 exhibited cell-cycle arrest at G2/M phase and proapoptotic activity [7].

Spectral analysis

Absorption spectra of the studied compounds were investigated in the general range from 200 to 400 nm. The maximum absorption wavelengths were found 254 nm for the studied compounds. Excitation wavelength was determined at the λmax 254 nm while the emission wavelengths were 362 nm for compounds 6–9, and 350 nm for 10 as shown in Table 3 and Fig. 2 for the starting agent 9-azidoacridine, and the triazoloacridinyl compounds 7–9 as representative examples. In the current work, complete resolution was obtained by using HPTLC plates, and each member of the studied compounds can be quantified at its specified RF value and using the specified wavelengths.

Table 3 Assigned wavelengths for UV and fluorescence measurementsFig. 2figure 2

Absorption (A) and emission (B) spectra of compounds 9-azidoacridine, 6‒9 in the range of 200‒500 nm

Mobile phase composition and saturation time

Due to the marked difference in polarities and functional groups for the studied drugs, we cannot use a single mobile phase for their separation. Thus, a mixture of solvents was used for separation of the studied compounds with different ratios of solvent components. Compact spots as well as good separation with reasonable RF values were obtained for each of the studied compounds by using the mobile phases indicated in Table 4 and Fig. 3 as representing example. Several trials were made by changing both the polarity and composition of different solvents. Chloroform or hexane was selected as one of the components of the mobile phase with acceptable resolution. However, the RF value was too small, so the solvent strength was increased by adding a polar solvent. Methanol was added to chloroform, while ethyl acetate was added to hexane to develop improved chromatograms. The mobile phase comprising chloroform‒methanol (8:2, V/V) showed good resolution. Saturation time ranging from 15 min up to 45 min was tried in order to obtain compact good separated spots for all investigated compounds. It was found that the highest optical intensities and good resolution were obtained at saturation time between 25 and 30 min. A 30 min saturation time was selected for all compounds in the subsequent work.

Table 4 Solvent systems used for separation of the studied compoundsFig. 3figure 3

Two-dimensional TLC densitogram for compounds 9-azidoacridine, 6, 7, and 8 at concentration level of 600 ng/band for all compounds and using mobile phase (chloroform‒methanol, 9:1, V/V)

Validation of the HPTLC method

The method was validated according to the International Council for Harmonisation (ICH) (QR2) guidelines and also complied with the United States Pharmacopeia (USP) 31-NF 26 for the determination of range, linearity, limits of detection and quantification, precision, selectivity, and specificity. For most of the calculations and statistical analysis, Excel 2016 was used.

Calibration range and linearity

According to the ICH guidelines, the range of analytical procedure is the interval between the upper and lower concentration of an analyte in the sample for which it has been demonstrated with a suitable level of precision, accuracy, and linearity, while linearity of an analytical procedure is its ability (within a given range) to obtain test results that are directly proportional to the concentration of analyte in the studied sample. Six concentration levels or more (three replicates for each) were selected to construct the calibration curves. The selected concentrations in the range from 10 to 500 ng/spot, the assigned wavelengths λmax 254 nm for the UV, and the wavelengths λex 254, λem 362, 350 nm for the fluorescence measurements were indicated for each compound (Table 5). Linear regression data revealed good linear relationships between the concentrations used and the corresponding peak areas with correlation coefficients ranging from 0.9913 to 0.9992. Figures 4, 5, 6, 7, 8, 910, 11, 12, and 13 show the linear calibration curves and the three-dimensional densitograms for calibration levels of the studied compounds 6–10.

Table 5 Statistical parameters of the regression equationsFig. 4figure 4

Three-dimensional HPTLC densitograms of the calibration ranges of compound 6 using UV (A) and fluorescence (B) measurements

Fig. 5figure 5

The linear calibration curves for the compound 6 measured spectrophotometrically (A) and spectrofluorometrically (B)

Fig. 6figure 6

Three-dimensional HPTLC densitograms of the calibration ranges of compound 7 using UV (A) and fluorescence (B) measurements

Fig. 7figure 7

The linear calibration curves for the compound 7 measured spectrophotometrically (A) and spectrofluorometrically (B)

Fig. 8figure 8

Three-dimensional HPTLC densitograms of the calibration ranges of compound 8 using UV (A) and fluorescence (B) measurements

Fig. 9figure 9

The linear calibration curves for the compound 8 measured spectrophotometrically (A) and spectrofluorometrically (B)

Fig. 10figure 10

Three-dimensional HPTLC densitograms of the calibration ranges of compound 9 using UV (A) and fluorescence (B) measurements

Fig. 11figure 11

The linear calibration curves for the compound 9 measured spectrophotometrically (A) and spectrofluorometrically (B)

Fig. 12figure 12

Three-dimensional HPTLC densitograms of the calibration ranges of compound 10 using UV (A) and fluorescence (B) measurements

Fig. 13figure 13

The linear calibration curves for the compound 10 measured spectrophotometrically (A) and spectrofluorometrically (B)

Sensitivity

The sensitivity of the method was determined in terms of limit of detection (LOD) and limit of quantitation (LOQ). LOD and LOQ were determined for the investigated compounds using the formula:

where K is a constant = 3.3 for LOD and 10 for LOQ, SDa is the standard deviation of the blank determination or standard deviation of intercept of the calibration curve, and b is its slope. The LOD values of the synthesized compounds 6–10 were found to be 11.02–51.09 ng/spot and 3.84–31.95 ng/spot for both spectrophotometric and spectrofluorometric method, respectively. These were the least amounts of analyte in a sample that can be detected but not necessarily quantitated as an exact value. The LOQ values for the synthesized compounds 6–10 were 33.39–154.82 ng/spot and 11.63–73.67 ng/spot for both spectrophotometric and spectrofluorometric method, respectively. These were the lowest concentrations of the drugs that were accurately detected and integrated by the instrument used.

The calculated LOD and LOQ values for all compounds studied are shown in Table 6, indicating high sensitivity of the proposed methods. Upon comparison of LOD and LOQ values obtained by the proposed method with those of the reported method for doxorubicin, there was a significant difference: LOQ was 12.5 ng/mL for doxorubicin and 11.63 ng/mL for compound 6 and LOD was 5 ng/mL for doxorubicin and 3.84 ng/mL for compound 6 indicating high sensitivity [14].

Table 6 Linearity ranges, limits of detection, and quantitation for the studied compounds 6‒10 determined by the proposed HPTLC methodSelectivity and specificity

Selectivity of analytical method is the measure of the extent to which the method can determine a particular compound in the analyzed matrices without interference from the sample matrix or degradation products. As indicated by the previous results for separation and determination of the studied compounds, there is no interference between them and the starting materials or intermediates peaks, and a good separation of their peaks was obtained in all cases. Figure 14 indicates the separation of compound 6 in the presence of their starting materials as representative example of the studied compounds. In addition, method selectivity was determined using peak purity test. The purity of the synthesized compounds was confirmed by 1H- and 13C-NMR spectroscopy, elemental analysis, and MS [7]. The peak purity of the studied compounds was assessed by comparing peaks of the studied compound and starting material at peak start, peak apex, and peak end position. The results of comparison indicated the closeness in these positions in both studied compound and starting material. The bands for all compounds were confirmed by comparing spectra, as shown in Fig. 15, peak area, and RF values of the sample spots. The method is considerably selective and suitable for the determination of the studied compounds in the pure form and in the presence of starting materials and intermediates.

Fig. 14figure 14

Two-dimensional TLC densitogram of compounds 6‒8 and starting material at concentration level of 600 ng/band for all compounds using the mobile phase chloroform‒methanol (9:1, V/V)

Fig. 15figure 15

Spectrum of compound 6 in comparison with spectra of starting materials as representative example of studied compounds with both UV (A) and fluorometric detection (B)

Precision

The precision of the method was expressed as relative standard deviations (RSD). RSD% values ranged from 0.34% to 4.47% indicating good repeatability and precision as shown in Table 7. The obtained precisions were satisfactory for quality control measurement.

Table 7 Precision studies of the developed method